Thinky Multi-Sensor
Thinky Multi-Sensor
Sensor for online measurement of temperature and speed for laboratory and pilot plant mixersThe Thinky Multi-Sensor is capable of measuring speed and temperature and recording the data using software. It connects to a PC via Bluetooth® and is powered by a button cell battery, which is designed to last for approximately 100 hours of operation.
Product Description
The Multi-Sensor is an accessory for our Thinky laboratory and pilot plant mixers. It is a special lid sensor, compatible with 150/300/550/650 ml HDPE beakers and the adapters that use the same lids, e.g. 310AD-W, 310AD-W200DSP-R2, 250AD-200DSP-BLOW2, 250AD-100DSP-MP, which all use the lid of the 300 ml version.
The sensor measures the speed and temperature up to five times per second. The latter is measured using a contactless infrared method.
Data is transferred from the sensor to the receiving PC via Bluetooth®. A corresponding receiver module for USB is supplied. The appropriate Windows software is also included in the scope of delivery. This allows all measurement data to be recorded and displayed in an Excel-compatible (.csv) table format.
The built-in CR2450 button cell supplies the sensor with power and is designed for an operating time of approx. 100 hours. The battery status is displayed within the software.
Downloads
Videos
Aktuell sind keine Videos verfügbar.
Specifications
Compatible mixers: Thinky ARE-250, SR-500, ARM-310, ARV-310, ARV-310P, ARV-310LED, ARE-500, ARV-501, ARV-930Twinwhen using the following mixing cups/adapters: 150ml HDPE cup, 300ml HDPE cup and adapters 310AD-W, 310AD-W200DSP-R2, 250AD-200DSP-BLOW2, 250AD-100DSP-MP, 550/650ml HDPE cup
Ambient conditions: 5–35 °C, 35–85% relative humidity (non-condensing)
Power supply: Panasonic CR2450
button cell
Bluetooth® receiver interface: USB
Transmission range: approx. 5 m
Measurement interval: 200 ms
Measurement method: IR thermometer, angular velocity measurement
Measuring range: 0–79.9 °C, 200–2,000 rpm (not compatible with degassing mode/defoaming mode)
Resolution accuracy: 0.1 °C, 1 rpm
Measurement error: ±1.0 °C, +/- 2% (rpm)
Unit weight: approx. 100/135/155 g, including 150/300/550 ml beaker
Dimensions: Beaker height + 27 mm
Software requirements: Windows 7, 8, 10 (32/64 bit)
Options and Accessories
• 24-month warranty•
Variants: 150ml cup, 300ml cup (also compatible with adapters 310AD-W, 310AD-W200DSP-R2, 250AD-200DSP-BLOW2, 250AD-100DSP-MP; others available on request), 550ml beaker (also compatible with 650ml beaker)