Gamry ParaCell – Measurement Cell Kit: Galvanic Corrosion and Electrochemical Noise
Gamry ParaCell – Measurement Cell Kit: Galvanic Corrosion and Electrochemical Noise
The measuring cell set for galvanic corrosion measurements and electrochemical noise enables the analysis of a wide range of flat specimens, either in relation to a graphite counter-electrode or a second working electrode in a symmetrical configuration.The Galvanic Corrosion and Electrochemical Noise Measurement Cell Set brings the working electrode and counter electrode into contact with one another via the 300 ml electrolyte capacity, ensuring that the electrode surfaces are identical. This fulfils the requirements for both electrochemical noise measurements and galvanic corrosion current measurements using a zero-resistance ammeter.
Product Description
For corrosion measurements using a potentiostat or galvanostat, the working electrode is attached to the polycarbonate end plate with the screw clamp of the base body via the contact body, in which a reference electrode is already screwed into the orthogonal Luggin capillary via the #11 Ace – Thred with the corresponding threaded bush.
A graphite block is attached to the opposite end plate as a counter electrode. Both end plates expose an electrode surface of 2.6 cm2 in contact with electrolytes, which can be additionally masked in the centre with an adhesive foil to avoid gap conditions at the working electrode.
For corrosion measurements using a zero resistance ammeter, a working electrode is attached to one end plate with the screw clamp of the base body via the contact body, while the second working electrode is attached to the opposite end plate. Both end plates expose an electrode surface of 2.6 cm2 in contact with electrolytes, both of which can be additionally masked in the centre with an adhesive film to avoid gap conditions. Ideally, the potential of the working electrode and the potential of the counter electrode can then be recorded simultaneously with the reference electrode due to the identical mixed potential of this galvanic chain.
Downloads
Videos
The Gamry Interface 1010 Potentiostat Starter PackSpecifications
DIMENSIONS30 cm (L) x 15 cm (W) x 23 cm (H) – with reference electrode
30 cm (L) x 15 cm (W) x 15 cm (H) – without reference electrode
SYSTEM
Material: Pyrex®, PTFE, Vycor glass, polycarbonate, ACE FETFE, graphite
Type:
Side plates for working and counter electrodes with centrally pressed-in cell body
Replaceable glass cell body with heating/cooling jacket
2 x side plates (2x #7 Ace – Thread)
Stainless steel frame with 4x threaded rods and 8x lock nuts
Volume: 250 ml – 400 ml
CONFIGURATION
Working electrode connection (type) Side panel clamping
device
Dimensions (min.) 25 mm x 25 mm
Working electrode material thickness (max.) 25 mm
Working electrode area (cm²) 2.6
Insulation (type) Foam rubber (side panel / contact)
Luggin capillary connection (type) #7 Ace – Thread
Luggin capillary / electrolyte contact Vycor® disc Ø, 3.5 mm
Reference electrode connection (type) #11 Ace – Thread
Reference electrode connection (Ø / mm) 9 – 11.5
Counter-electrode connection (type) Clamping device, side panel
Dimensions (min.) 25 mm x 25 mm
Counter-electrode material thickness (max.) 25 mm
Counter-electrode dimensions 25 mm x 25 mm x 5 mm (graphite)
Working/counter-electrode area (cm²) 2.6
Insulation (type) Foam rubber (side panel / contacts)
Additional connections (type) 3x #7 Ace – Thread for inert gas purging, electrolyte volume exchange or sampling
Options and Accessories
• Interface 1000 Potentiostat/Galvanostat/ZRA
• Interface 1010 Potentiostat/Galvanostat/ZRA
• Reference 600 Potentiostat/Galvanostat/ZRA
• Reference 600+ Potentiostat/Galvanostat/ZRA
• Reference 3000 (AE) Potentiostat/Galvanostat/ZRA
• ECM8
• ParaCell end plates, polycarbonate
• ParaCell Cell Body, Polycarbonate
• End brackets, stainless steel
• Reference Electrode Bridge Tube, Pyrex
• #11 Ace-thread bushing, nylon with O-ring
• Sample brackets with banana jacks, stainless steel
• Graphite block, counter electrode
• #7 Ace-thread bushings, nylon with O-ring
• #7 Ace-thread plugs, nylon
• PortHoles electrochemical sample masks, 3×3″, 1 cm², 60 per pack
• Calomel (SCE) reference electrode
• Silver/silver chloride reference electrode
• Mercury sulphate reference electrode (without chloride anions)
• Mercury oxide reference electrode for alkaline media
• Silver/silver nitrate reference electrode for non-aqueous electrolytes
• SCE refill solution (120 ml)
• Ag/AgCl refill medium (120 ml)
• Vycor frit with Teflon heat-shrink tubing, 3.5 mm (5 pieces)
• KT glass frit with Teflon heat-shrink tubing, 3.5 mm (5 pieces)
• Polymer exchange frits for mercury oxide reference electrode
Applications
• Two, Three and Four Electrode Experiments
• Understanding iR Compensation
• THE Method for Crevice Corrosion Repassivation Potentials
• Rapid Electrochemical Assessment of Paint
• Polarisation Resistance Tutorial – Getting Started
• Comparison of Corrosion Rate Calculated by EFM, LPR, and EIS
• Getting Started with Electrochemical Corrosion Measurement
• EIS of Organic Coatings and Paints
• OptiEIS™: A Multisine Implementation