Hot Disk TPS 500
Hot Disk TPS 500
The basic instrument for the quick and easy measurement of thermal conductivity, temperature conductivity and heat capacity.The compact TPS 500 measuring instrument offers an affordable entry point into thermal conductivity measurement. Even in its basic configuration, it allows for the simultaneous measurement of thermal conductivity, thermal resistance and specific heat capacity (per unit volume).
Product Description
The Hot Disk TPS 500 determines the thermal conductivity, the thermal diffusivity and the heat capacity (volumetric) in one measurement, typical measuring times are 20 seconds. The device covers a measuring range from 0.03 to 100 W/mK.
The hot disc technique is also known as the “Transient Plane Source” (TPS) technique. The method works dynamically in a transient manner and is therefore very fast. The flat sensor serves as a heat source and temperature sensor at the same time. For standard measurements with the TPS 500, it is placed between two samples of identical material. The samples only need to have a flat surface, but may have rough or porous surfaces. The effort required for sample preparation is therefore minimal.
In addition to solid (bulk) materials, powders, granulates, pastes or viscous liquids can also be measured with the system.
In addition to the standard software, the TPS 500 also includes single-side technology – this enables measurements to be taken when only one sample piece is available.
Downloads
Videos
Aktuell sind keine Videos verfügbar.
Specifications
Thermal conductivity measurement range:0.03 to 100 W/mK (standard for isotropic samples)
Thermal diffusivity measurement range:
0.02 to 40 mm²/s (standard for isotropic samples)
Heat capacity measurement range:
0.1 to 4.5 MJ/m³K
Temperature range:
Room temperature or -100
to 200°C with suitable temperature control
Reproducibility
:
2% (thermal conductivity)
10% (thermal conductivity, sensor radius 6.4 mm)
12% (heat capacity, sensor radius 6.4 mm)
Measurement
accuracy:
better than 5% (thermal conductivity)
Measurement time:
2.5 to 2560 seconds
Minimum sample geometry:
3 mm (thickness) x 13 mm (diameter or square) for bulk materials
Maximum sample geometry:
unlimited
Compatible sensors:
Kapton: #5465, #5501
Mica: #5465, #5501
Options and Accessories
• Temperature
control equipment• Special sample holders for powders and liquids